Titan Fixed Benzene Detector

Description

Set to change the game in refinery applications, Titan is the first truly selective benzene monitor on the market. Titan is a fixed, continuous, real-time, benzene-specific monitor designed for petrochemical environments, providing the ultimate device for plant and worker safety.

Features
Best available benzene detection
Ultimate safety
Ease of use & service
Ease of use & service

Description

Set to change the game in refinery applications, Titan is the first truly selective benzene monitor on the market.

 

Titan is a fixed, continuous, real-time, benzene-specific monitor designed for petrochemical environments, providing the ultimate device for plant and worker safety.

 

The instrument is wall-mounted with a dynamic detection range or 0 – 20 ppm with 0.1 ppm sensitivity to benzene. Titan samples gas from the environment once per minute and within just 60 seconds an accurate benzene measurement is displayed.

 

Titan provides an immediate warning alarm system with two operator configurable levels, ensuring workers are kept safe and protected to the standards required on site. The instrument incorporates two relay outputs allowing you to install your own required alarm system.

 

Titan’s continuous, real-time measurement allows trends to be monitored over time, and communicated via 4-20 mA or RS485. Data is stored internally and can be downloaded remotely for analysis.

 

Designed to be easily installed and serviced, the modular design allows the Ex d case to be installed well in advance of commissioning. The internal service module can beremoved and replaced as a remotely serviceable plug and play cartridge.

 

Titan incorporates Ion Science patented photoionisation detection (PID) sensor technology with proven resistance to humidity and contamination, ensuring optimal performance, however challenging your environment.

 

 

Features

 

Best available benzene detection

·         Fast and accurate detection of benzene down to 0.1 ppm

·         Minute by minute sampling providing continuous real-time data

·         Robust separation method ensures benzene-specific readings

·         Internally regulated heating for stable operation at extreme temperatures

Ultimate safety

·         Clear display & visual alarms for indication of benzene levels

·         Two user-definable independent alarm levels including optional real-time STEL calculation

·         Two relay outputs provide an immediate warning alarm

·         Meets ATEX & IECE, UL and CSA standards

Ease of use & service

·         Modular design for ease of installation and servicing

·         Service-free six months operation

·         Simple 2-button interface for menu navigation

·         2-year warranty when instrument registered online

Best available PID sensor technology

·         Patented PID sensor technology

·         PID sensor independently verified as best performing on the market

·         In-built humidity resistance with no need to compensate

·         Anti-contamination design for extended field operation

 

 

Specifications

Sensor Type

PID, 10.6 eV lamp coupled with selective filtering

Selectivity

Benzene specific within typical petrochemical matrix

Range

0 – 20 ppm

Measuring Frequency

One minute

Lower Limit of Detection

0.1 ppm

Accuracy

±0.05 ppm or ±10%, whichever is greater

Temperature Stability

Internally heated

Sample Flow

200 ml/min

Display

Graphical BW LCD 64 x 128 pixels with back light
Bright LEDs for normal operation, fault and alarm
Two magnetic switches for menu operation

Hazardous Area Safety Approvals

ATEX/IECEx: II 2G Ex d II B+H2 T4 Gb (-20 oC to +55 oC)

Ingress Protection Rating

Designed to IP65 (cable gland dependant)

Humidity

0 – 100% RH

Minimum Service Period

Selective filter six months
PID lamp one year

Cable Entry

Two 3/4 NPT gland threads

Power

Vin (nom) 24 V DC @ 3.2 A (recommended) Vin (max) 32 V DC @ 2.4 A
Vin (min) 19 V DC @ 4.0 A

Communication

Isolated 4-20 mA output
RS 485 Modbus
USB when lid removed with PC software

Data Storage

On board MMC, minimum of six months

Relay

Two isolated outputs, voltage free, 24 V DC @ 1.25 A max

Weight and Dimensions

15 kg (33 lbs)
219 x 219 x 172 mm (8.6 x 8.6 x 6.8” approx.)

 Gas Sample Line

10 m max length, PTFE, 6 mm OD, 4 mm ID