SF6 AREACHECK P2 FIXED SF6 MONITOR

SF6 AREACHECK P2 FIXED SF6 MONITOR

Description

DESIGNED FOR CONTINUOUS SF6 AREA MONITORING, THE AREACHECK P2 RAPIDLY DETECTS LOW-LEVEL SF6 LEAKS PROTECTING WORKERS AND THE ENVIRONMENT.

Features
Award winning Negative Ion Capture (NIC) technology
Fast, accurate detection down to highly sensitive levels
No cross sensitivity with other gases or moisture in air
Safety

Description

Designed for fixed continuous SF6 area monitoring, the AreaCheck P2 rapidly detects low-level SF6 leaks. AreaCheck P2 utilises relay output for immediate awareness of detected leaks, and presents no cross sensitivity with any other gas or moisture in the air. 

Instrument setup is carried out easily via a computer or network interface, allowing optimal protection from unauthorised access. The SF6 AreaCheck P2 has a built-in low flow alarm and self-diagnostic testing. 

The instrument’s serviceable components are comprised in the user exchangeable SmartSensor, with a lifetime of up to 12 months. Measuring stations are maintenance-free, ensuring instrument downtime is minimised. 

AreaCheck P2 has no filter problems due to its minimum air intake (compared to pump operated systems). 

Can be used on an RS485 network, it also outputs RS232 and has relay outputs for standalone functionality.

 

 

Features

Best proven SF6 leak detection 
  • Award winning Negative Ion Capture (NIC) technology 
  • Fast, accurate detection down to highly sensitive levels 
  • No cross sensitivity with other gases or moisture in air 

Safety 
  • Relay output for immediate awareness of leaks detected 
  • Audio and visual alarms provide clear indication of SF6 leaks 
  • Complete safe set up via PC or the Network Interface protects from unauthorised access 
  • Built in low flow alarm and diagnostics test

Flexibility 
  • Can be used on an RS485 network, it also outputs RS232 and has relay outputs for standalone functionality 
  • Exchangeable SmartSensor (12 month lifetime) 
  • Maintenance free operation 

 

 

Specifications

DETECTION PRINCIPLE: SF6: NIC

RANGE: 0 - 2000 ppm SF6

RESOLUTION: 500 ppm SF6 /10ppm

MAINS POWER: 100 - 240 VAC, 50/60 Hz

POWER CONSUMPTION: 18VA

OPERATING TEMPERATURE:  -5°C to 45°C

STORAGE TEMPERATURE: -20°C to 60°C

OPERATING HUMIDITY RANGE: 10 - 90% non condensing

MAX LOAD, RELAY OUTPUT: 2,5A / 230 VAC

SIZE: H 280 x B 165 x T 125 mm

PROTECTION CLASS: IP 52

NOISE LEVEL OF AUDIBLE ALARM: > 75 dbA, 1m

WEIGHT: 1,5 Kg (w/o wall mounting bracket)

FUSE: T 1A (Slow Blow)

 

 

Application
 

  • SF6 leak testing and measurement in high voltage switchgear (GIS) 
  • Leak integrity tesing on medical, refrigeration and air conditioning equipment containing SF6 and (H)CFCs 
  • Breathing apparatus testing 
  • Medical device testing